Beijing Aerospace Changfeng Co.,Ltd Medical Devices Branch
|
|
| |
|
|
|
|

|

|
Click to enlarge image
|
Model Number |
ACM603 |
Anesthesia Ventilator
Ventilation modes: IPPV, Manual
Pneumatically driven and electrically controlled
Time switch, volume control
Parameter Settings
Tidal volume: 50ml‐1200ml, manually monitoring tidal volume and respiratory rate
Respiratory rate: 6bpm~40bpm
I/E Ratio: 1:1, 2:3,1:2, 1:3
Monitoring Parameter
Tidal volume
Airway pressure
Manual mode monitoring
Alarms Parameter
Upper airway pressure limit: 2kPa~6kPa
Lower airway pressure limit: 0.5kPa~2kPa
Pressure limit for safety valve: 5.1kPa~6KPa
Silence time for alarm: 120seconds
Anesthesia Machine
Flowmeter:
O2/N2O four tube flowmeter
Low flowrate: 0.1L/min~2L/min
High flowrate: 2.5L/min~10L/min
ORC device
The proportion between O2 and N2O can be continuously adjustable to assure the O2 concentration above 25%.
Safety Performance:
Alarm for oxygen supply failure
Cut‐off of N2O
Spill valve
Safety valve: 5.1kPa~6KPa
O2 Flush:
35L/min~75L/min
Gas Supplies:
O2: 0.3Mpa~0.4Mpa
N2O: 0.3Mpa‐0.4Mpa
Vaporizer
The standard configuration is one ACM102 vaporizer, which uses enflurance or isoflurane as the anesthetic agent. Enflurane�s concentration adjusting range is 0.5%~5%VOL. Isoflurane�s concentration adjusting range is 0.5%~4%VOL.
It can be equipped with double vaporizers, which can be interlockable. It also possesses the function of flow and temperature compensation.
Writing Board
Pull‐Push Style, easy for operation
Oxygen Inhaler (Optional)
Easy but precise equipment.
|
|
|
|
| |
|
|
|